| RSLCnano LC Systems Specifications |
|---|
| Key Pump Specifications |
| Flow Rate Ranges: | HPG Nano Pump: 20 nL/min–50 μL/min using precalibrated flow selector LPG Micro Pump: 10–2500 μL/min |
| Pressure Range: | 2–80 MPa (300–11,600 psi) 2–50 MPa (300–7250 psi), loading pump |
| Number of Solvent Channels: | HPG Nano Pump: 2 LPG Micro Pump: 3 |
| Flow Calibration: | Semi-automated |
| Proportioning Accuracy: | <1% of full scale |
| Proportioning Precision: | Typically <0.1% SD |
| Retention Time RSD in Gradient Mode at 300 nL/min: | <0.2% RSD or <0.1 min SD, whichever is greater |
| Gradient Delay Volume: | <25 nL (pump) and <350 nL (system in preconcentration configuration) |
| Eluent Bottles: | 2 × 100 mL 3 × 500 mL |
| Solvent Degassing: | External (optional) |
| Wetted Parts: | Binary Nano Pump: SST, titanium, PEEK™, UHMW-PE, PTFE, FEP, ruby, sapphire, ZrO2, Al2O3, fused silica Ternary Loading Pump:ST, titanium, PEEK, UHMW-PE, PTFE, FEP, ruby, sapphire, ZrO2, Al2O3, Kalrez® |
Dimensions NCS-3500 (h × w × d): | 36 × 42 × 51 cm (14.1 × 16.5 × 20 in.) |
| Weight: | 32 kg (70.6 lb) |
| Power Requirements: | 100–120 V, 60 Hz 200–240 V, 50 Hz; max 300 VA |
| PC Connection: | USB 2.0; USB hub with three integrated sockets |
| I/O Interfaces: | Two digital inputs and two programmable outputs |
| Additional Communication Port: | 15-pin D-Sub port for connection of a solvent rack or degasser |
| GLP Features: | System wellness monitoring, column tracking |
| Safety Features: | Leak sensor, active rear-seal wash system, excess pressure monitoring |
| Key Column Compartment Specifications |
| Temperature Range: | RT + 10 ˚C to 75 ˚C |
| Temperature Accuracy: | ± 0.5 ˚C (at 50 ˚C setpoint) |
| Temperature Stability: | ± 0.1 ˚C (at 50 ˚C setpoint) |
| Heat-Up Time: | From 30 ˚C to 60 ˚C in 15 min |
| Switching Valves: | Up to two 6-port or 10-port, 2-position low-dispersion valves Port-to-port volume: 124 nL Maximum pressure: 100 MPa (14,500 psi) |
| Column Capacity: | Up to 3 columns Up to 100 cm length (75 μm i.d., coiled) |
| Safety Features: | Humidity sensor, leak sensor |
| Key Autosampler Specifications |
| Injection Volume Range: | 10 nL–125 μL |
| Sample Formats: | 96 (deep) well plate, 384 (deep) well plate, sealed or open; 40 standard autosampler vials, 1.8 mL, sealed or open |
| Sample Capacity: | 3× well plate (128 × 86 mm) 15× 10 mL vials for reagents, diluents, and transport liquids |
| Injection Cycle Time: | 30 s for a 1 μL full loop injection |
| Injection Methods: | Full loop, partial loop, low-dispersion injection Microliter pick-up, user-defined injection programs |
| Injection Technique: | Needle-in-needle with programmable needle wash |
| Injection Valve Precision: | <0.4% RSD for 1 μL full loop injection |
| Injection Linearity: | Corr. coeff. > 0.9995 from 100–500 nL injections |
| Carryover: | <0.02% with needle wash (caffeine) |
| Optional Sample Cooling | 4 ˚C–45 ˚C, at least 22 ˚C below ambient |
| Fraction Collection: | Micro fraction collection option [up to 34.5 MPa (5000 psi)] |
| Wetted Parts: | PEEK, Stainless steel, PCTFE, fused silica |
Dimensions (h × w × d): | 36 × 42 × 51 cm (16 × 16.5 × 20 in.) |
| Weight: | 22.7 kg (50 lb) including cooling |
| Power Requirements: | 100–120 V, 60 Hz 200–240 V, 50 Hz |
| PC Connection: | USB; USB hub with three integrated sockets |
| I/O Interfaces: | Four digital inputs and four programmable outputs |
| Key Detector Specifications |
| Data Collection Rate: | Up to 200 Hz (in single wavelength mode) |
| Maximum Number of Channels: | Four |
| Drift: | 4.0 mAU/h |
| Wavelength Range: | 190–900 nm ± 1 nm |
| Noise: | Typically <0.05 mAU at 254 nm |
| Lamp: | Deuterium lamp, Tungsten lamp |
| Flow Cell Volume: | 3 nL for nano LC 45 nL for capillary LC 180 nL for micro LC |
Dimensions (h × w × d): | 16 × 42 × 51 cm (6.3 × 16.5 × 20 in.) |
| Weight: | 15 kg (33 lb) |
| Power Requirements: | 100–120 V, 60 Hz 200–240 V, 50 Hz |
| PC Connection: | USB |
| I/O Interfaces: | Two digital inputs and two programmable outputs Two analog inputs available as an option via DAC plug-in module |