The Thermo Scientific Dionex UltiMate™ 3000 Dual-Gradient Analytical Pump DGP-3600SD is the pump of choice for advanced analytical chromatographic techniques. It supports all standard LC applications and UHPLC applications up to 62 MPa (9000 psi) and is ideal for increasing sample throughput, achieving higher chromatographic resolution, or automating sample preparation steps such as analyte enrichment or matrix elimination.
- Serial dual-piston operating principle, low-pressure gradient proportioning
- Two ternary Rapid Separation pumps in a single instrument enclosure
- Wide operating flow rate range from 200 μL/min to 10 mL/min with high flow rate accuracy
- Support of operating pressures of up to 620 bar (9000 psi)
- SmartFlow™ technology automatically ensures optimal chromatographic performance at any operation condition
- Automatic compressibility compensation, no manual solvent setting needed
- Unique SpinFlow™ mixing design perfectly balance gradient delay volume against mixing efficiency
- Extensive mixer portfolio (35–1550 μL) to cover all application needs
- Active rear-seal wash for maximum seal lifetime, and reduced instrument down time
- Easy control of x2 dual systems with Chromeleon™ software including control of two gradients in parallel
The DGP-3600SD is recommended for research and other analytical laboratories that require a general purpose LC pump capable of standard as well as advanced two-dimensional techniques.
Performance Examples
UltiMate® 3000 HPLC systems deliver improvements in HPLC performance while ensuring easy, reliable operation.The UltiMate 3000 series operates over a wide range of flow rates, including preparative, analytical, micro, and nano/capillary scales, supporting applications such as discovery processes, quality control, and advanced research like proteomics.
Rapid Separation LC offers the highest flow-pressure range to meet your chromatographic goals whatever the may be. Developing separations in seconds rather than in minutes requires a system that provides excellent precision, speed, and performance. Use the RSLC Method Transfer Calculator for quick and easy conversion of conventional HPLC to UHPLC methods.
Optimum Mixing of Solvents with the UltiMate RSLC Pump

Ultrafast gradient separations require a small gradient delay volume to ensure the gradient is correctly applied to the column. However, high mixing volumes are typically required to ensure the mobile phase components are completely mixed.This is particularly important when working with UV detectors, as incomplete mixing may lead to ripple effects in the chromatography baseline.
With very fast separations even small fluctuations in the mobile phase composition can compromise the precision of the peak retention time. Therefore, complete mixing is necessary to achieve a constant, flat baseline, high signal-to-noise ratios, and accurate peak identification.
Different types of applications have different requirements regarding mixing efficiency. Some require almost no mixing, while others require additional mixing volume to achieve a smooth baseline. A water/acetonitrile gradient using TFA as a modifier is an example of this type of gradient that requires precise mixing efficiency
The new proprietary mixing concept from Dionex addresses all these needs with:
- A two-step process of radial mixing followed by longitudinal mixing to ensure the highest possible mixing efficiency.
- Mixing volume that can be easily adapted to fit specific method requirements. The default mixing volume (400 µL) can be easily changed using mixing kits of 35, 200, 800, or 1500 µL (effective mixing volume).
- Available for the UltiMate® 3000 Binary, Quaternary, and Dual-Gradient RSLC pumps providing ultrahigh resolution and ultrafast separations.
| Dual-Gradient Analytical Pump Specifications |
|---|
| Operating principle | Serial dual-piston |
| Gradient formation | Dual low-pressure gradient proportioning |
Flow range (settable range) | 0.2–10 mL/min (0.001–10 mL/min) |
| Flow accuracy | ±0.1% |
| Flow precision | < 0.05% RSD or < 0.01 min SD, whichever is greater |
| Pressure range | 2–62 MPa (290–9,000 psi) |
| Pulsation | Typically <1% or <0.2 MPa, whichever is greater |
| Proportioning accuracy | ±0.5% (of fullscale) |
| Proportioning precision | < 0.15% SD @ 2mL/min |
| No of eluent lines | 2 × 3 |
| Gradient delay volume | 690 μL by default |
| Solvent degassing | External (optional) |
| Dimensions (h × w × d) | 16 × 42 × 51 cm (6.3 × 16.5 × 20 in) |
| Weight | 16.5 kg (36.3 lb) |
| PC connection/communication | USB 2.0 for PC connection; USB hub with 3 sockets integrated |
| Hardware |
| DGP-3600SD - UltiMate 3000 Dual-Gradient Analytical Pump |
5040.0061 |
| SRD-3600 UltiMate 3000 Integrated Solvent and Degasser Rack, 6 Channels |
5035.9230 |
| Accessories |
| Static Mixer Kit, Micro, 35 µL Mixing Volume |
6040.5000 |
| UltiMate 3000 Pump Diagnostics Kit |
6035.3000 |