Home >
Dionex Products >
Ion Chromatography >
IC and Reagent-Free IC Modules >
IC and RFIC Column Compartments >
ICS-5000 DC Detector/Chromatography Compartment
ICS-5000 DC Detector/Chromatography Compartment
This compartment houses the detector, suppressor control, column thermal controller, and the preparation module. This combination provides complete tubing thermal stability and short connections between components. The CD and ED detectors are easy to install for immediate use.
- High performance dual zone thermal control provides two independent temperatures
- Innovative electrochemical detection technology
- CD conductivity detector features advanced electronics for superior detection
- Dual conductivity and electrochemical detection provides increased flexibility and utility
- Optional AM Automation Manager with flexible valving configurations for eluent addition and sample handling
- Supports AutoPrep, preconcentration, AutoNeutralization, matrix elimination, and postcolumn reagent addition
- Valve control integral to software, eliminating complicated TTL or relay programming
With the DC's flexible configuration, dual zone thermostatting, and Chromeleon® software control, your possibilities have more than doubled. Configure your system for simultaneous injection and perform both anion and cation analysis with one injection. Perform two different anion applications or eliminate dilutions from unknown samples by running the same application with large loop/small loop injections.
| ICS-5000 DC Detector/Chromatography Compartment Specifications |
|---|
| DC models: | Dual Temperature Model Only |
Dual Temperature Model: independent temperature control | Upper Section Temp. Range: 10–40 °C Lower Section Temp. Range: 10–70 °C Temperature Accuraccy: ± 0.5 °C Temperature Stability: ± 0.2 °C Temperature Precision: ± 0.2 °C |
| Lower Zone: (Analytical) | Injection valves: 1 or 2 ea., 6 or 10 port Up to two column sets, 1–9 mm Maximum column length: 250 mm Precolumn heat exchangers: 2 |
| Cap IC Cube: (Optional - but required for capillary chemistries) | 1 or 2 IC Cube module hold the following: Injection Valves: Up to 2 (one per cube) high pressure valves, 4-port, 2 position EG Degasser Column and Guard Carbonate Removal Device Capillary Suppressor (See Suppressor section) IC Cubes and cartridges are customer installable with preformed tubing and color-coded labeling Capillary column heater 15 to 85 °C |
| Automation Manager: (Optional) | Injection valves: Up to 2 ea., high pressure valves, either 6 or 10 port Low pressure valves: Up to 2 ea., Teflon®, 2 or 3 way Reaction Coil Heater: Holds up to two reaction coils Temperature Range: 5 °C above DC, up to 80 °C |
| Leak detection: | Optical leak sensor |
| Power Requirements: | 90–265 V ac, 47–63 Hz |
Dimensions (h × w × d): | 42 × 44.5 × 57 cm (16.5 × 17.5 × 22.5 in.) |