| | | | IonPac AS5 Anion-Exchange Column |
| The IonPac® AS5 is a low-capacity, hydroxide-selective, anion-exchange column designed for the separation of higher-valence anions, including polyphosphates, oxyanions, EDTA complexes, metal cyanide complexes, and hydrophobic anions such as iodide, thiosulfate, and thiocyanate. The AS16 column is recommended for hydrophobic anions and polyphosphates. Use the AS5 column for metal-EDTA complexes. Use the AS5 column for cyano-transition metal complexes. The AS5 can be used for hydrophobic anions, including iodide, thiocyanate, and thiosulfate. The high-capacity AS16 or AS20 columns are recommended for hydrophobic anions and highly charged anions.
| | IonPac AS5 Anion-Exchange Column Specifications | | Dimensions | IonPac AS5 Analytical Column: 4 × 250 mm IonPac AG5 Guard Column: 4 × 50 mm | | Maximum operating pressure |
2000 psi
| | Mobile phase compatibility |
pH 0–14; 0–3% HPLC solvents
| | Substrate characteristics | Bead Diameter: 15 µm Cross-Linking (%DVB): 2% | | Latex characteristics | Functional Group: Alkanol quaternary ammonium ion Latex Cross-Linking: 1% Latex Diameter: 120 nm Hydrophobicity: Low | | Capacity |
20 µeq (4 × 250 mm analytical column)
| | Column construction |
PEEK with 10-32 threaded ferrule-style end fittings. All components are nonmetallic.
|
| | Order Information | | IonPac AS5 Analytical Column (4 x 250 mm) | 035395 | | IonPac AG5 Guard Column (4 x 50 mm) | 035396 |
| |
|