Specifications
 
Printer Friendly

Specifications

Specifications

ICS-900 Entry-Level IC System Specifications

Pump

Type

Serial dual-reciprocating piston, microprocessor-controlled constant stroke, variable speed

Construction

Chemically inert, metal-free PEEK pump heads and flow paths compatible with aqueous eluents of pH 0–14 and reversed-phase solvents

Pump Operating Pressure

0–35 MPa (0–5000 psi)

Flow Rate Range

0.00 mL/min to 5.00 mL/min with settable flow increment at 0.01 mL/min without changing pump heads

Flow Precision

<0.1%

Flow Accuracy

<0.1%

Pressure Ripple

<1%

Eluent Bottle Pressure

None required

Vacuum Degasser

Optional with external vacuum degasser

Gradient Capabilities

Optional with RFC-30

Eluent Generation

Optional with RFC-30

Injection Valve

6-port, 2-position Rheodyne valve, electrically activated

Column Heater/ Thermostat

Optional External Heater CTS-10,
25 to 85 ˚C, minimum 5 ˚C above ambient

Suppressors and Control

Chemical Suppression

Yes, anion and cation, 2 and 4 mm

Displacement Chemical Regeneration

Yes, anion and cation, 2 and 4 mm

Electrolytic Suppression, Self-Regenerating

Yes with RFC-10

Electrolytic Suppression, Self-Regenerating with External Water

Yes with RFC-10

Suppression Capacity

Anion SRS 300 (4 mm): 200 µeq/min
Cation SRS 300 (4 mm): 110 µeq/min
Anion SRS 300 (2 mm): 50 µeq/min
Cation SRS 300 (2 mm): 37.5 µeq/min
Anion MMS 300 (4 mm): 150 µeq/min
Cation MMS 300 (4 mm): 150 µeq/min
Anion MMS 300 (2 mm): 37.5 µeq/min
Cation MMS 300 (2 mm): 37.5 µeq/min
Anion AES: 25 µeq/min
Cation AES: 25 µeq/min
AMMS-ICE : N/A

Void Volume

SRS 300 (4 mm): <50 µL
SRS 300 (2 mm): <15 µL
MMS 300 (4 mm): <50 µL
MMS 300 (2 mm): <15 µL
AMMS-ICE 300 (4 mm): <50 µL
AMMS-ICE 300 (2 mm): <15 µL
Anion AES: <35 µL
Cation AES: <35 µL

Sequential Suppression for Anions

Yes, MMS + CRD 300

Suppressor Wear Parts

None; peristaltic pump and inline filters not required

Detector

Conductivity Detector

Bipolar heated conductivity cell, microprocessor-controlled digital signal processing over the range 0–10,000 µS/cm

Detector Linearity

1%

Maximum Cell Pressure

10 MPa (1500 psi)

Resolution

0.0047 nS/cm

Cell Calibration

Factory calibrated as a new system and calibrated in the field individually with memory retention and transfer

Flow Cell Chemical Compatibility (standard)

Fully compatible with all typical Ion Chromatography eluents including MSA

AutoSampler

Automation Using Autosampler

Yes, Dionex AS40, AS-HV, AS Autosamplers or third party

Sequential/Simultaneous Injection

Yes, depending on autosampler capabilities

Automated Dilution

Yes, available with AS Autosampler

Dilution Factor, AS Autosampler

1:1 to 1:1000

Dilution Time, AS Autosampler

15 seconds with sample overlap

Inline Sample Degassing

Optional with CRD 300/200

Inline Filtration

AS40 Autosampler or inline filter

Automation Flexibility

High, conditionals using Chromeleon softwareand post run features

Semiautomated System Qualification

Yes, IQ/OQ/PQ

System Software

Software

Chromeleon Chromatography Management Software, requires Microsoft Windows® 2000, XP, or Vista

Automated Procedure Wizards

Yes, standard feature

System Wellness and Predictive Performance

Yes, standard feature

Virtual Column Simulator

Yes; evaluation mode standard,

isocratic and gradient optional

Application Templates

Yes, standard feature

Multivendor Automation Support of 3rd Party

Yes, fully controls over 300 different instruments from more than 30 manufacturers, including GC, HPLC, and MS

Customizable System Control Panels

Yes, standard feature

System Status Virtual Channels

Yes, standard feature

Power Failure Protection

Yes, standard feature

Sequential Injection

Yes, standard feature for independent dual system analysis

System Trigger Commands and Conditionals

Yes, standard feature

Daily Audit Trail

Yes, standard feature

Sample Audit Trail

Yes, standard feature

Multiple Network Control and Network Failure Protection

Yes, optional

System Calibration Storage

Yes, factory, present and previous. Completely user selectable

Customized Reporting

Yes, standard feature with unlimited report workbooks

Physical Specifications

Dimensions
(h × w × d)

33 × 24 × 40 cm
(13 × 9.5 × 15.75 in.)

Weight

10 kg (22 lb)

Power Requirements

100–240 V ac, 50–60 Hz, autoranging

System Flow Path

All polymeric (PEEK), anion or cation configurations

ICS-1000 Basic Integrated IC System
 

Chromeleon Chromatography Management System
Whether you use HPLC, IC, or GC—whether your scope is a single instrument, a laboratory, a department, or an enterprise—Chromeleon adapts to your needs. 
AS40 Autosampler

Automate sample loading with the economical AS40.

     
  Specifications  
 
Order Literature   Privacy   Feedback   Terms of Use   Trademarks   Site Map
© 2008 Dionex Corporation. All rights reserved. www.dionex.com